General Specifications |
Display |
73.5 mm x 104 mm |
Battery |
8 AA alkaline cells, IEC LR6 |
Dimensions (H x W x L) |
184 mm x 211 mm x 93 mm |
Weight |
1.3 kg |
Temperature |
Operating temperature |
-10 °C to +50 °C |
Storage temperature |
-20 °C to +60 °C |
Operating humidity |
Non-condensing (<10 °C) |
|
≤80 % RH (10 °C to 30 °C) |
|
≤50 % RH (30 °C to 50 °C) |
Altitude |
Operating altitude |
2000 m |
Storage altitude |
12000 m |
Overvoltage category |
CAT IV 600 V |
Safety |
|
General |
IEC 61010-1, pollution degree 2 |
|
IEC 61010-2-033:CAT IV 600 V |
|
IEC 61557-1 |
Insulation resistance measurement |
IEC 61557-2 |
Effectiveness of protective measures |
IEC 61557-16 |
IP Rating |
IEC 60529 IP40 |
Electromagnetic Compatibility (EMC) |
International |
IEC 61326-1: Electromagnetic environment for portable equipments |
|
IEC 61326-2-2 CISPR 11: Group 1, Class A |
Group 1: Equipment has intentionally generated and/or uses conductively-coupled radio frequency energy that is necessary for the internal function of the equipment itself.
Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and radiated disturbances.
Caution: This equipment is not intended for use in residential environments and may not provide adequate protection to radio reception in such environments.
Electrical Specifications |
The meter's accuracy is specified for one year after calibration at operating temperatures of 10°C to 30°C. For operating temperatures outside the range (-10°C to +10°C and +30°C to +50°C), add ± 0.25% per °C, except on the 20% bands, add ± 1% per °C. |
Insulation resistance testing |
Test voltage |
Range |
Resolution |
Accuracy |
250 V |
<200 kΩ |
Unspecified |
Unspecified |
200 kΩ to 500 kΩ |
1 kΩ |
0.05 |
0.50 MΩ to 5.00 MΩ |
10 kΩ |
0.05 |
5.0 MΩ to 50.0 MΩ |
100 kΩ |
0.05 |
50 MΩ to 500 MΩ |
1 MΩ |
0.05 |
0.50 GΩ to 5.00 GΩ |
10 MΩ |
0.05 |
5.0 GΩ to 50.0 GΩ |
500 MΩ |
0.2 |
>50 GΩ |
Unspecified |
Unspecified |
500 V |
<200 kΩ |
Unspecified |
Unspecified |
200 kΩ to 500 kΩ |
1 kΩ |
0.05 |
0.50 MΩ to 5.00 MΩ |
10 kΩ |
0.05 |
5.0 MΩ to 50.0 MΩ |
100 kΩ |
0.05 |
50 MΩ to 500 MΩ |
1 MΩ |
0.05 |
0.50 GΩ to 5.00 GΩ |
10 MΩ |
0.05 |
5.0 GΩ to 10.0 GΩ |
100 MΩ |
0.05 |
10.0 GΩ to 50.0 GΩ |
500 MΩ |
0.2 |
50 GΩ to 100 GΩ |
5 GΩ |
0.2 |
>100 GΩ |
Unspecified |
Unspecified |
1000 V |
<200 kΩ |
Unspecified |
Unspecified |
200 kΩ to 500 kΩ |
1 kΩ |
0.05 |
0.50 MΩ to 5.00 MΩ |
10 kΩ |
0.05 |
5.0 MΩ to 50.0 MΩ |
100 kΩ |
0.05 |
50 MΩ to 500 MΩ |
1 MΩ |
0.05 |
0.50 GΩ to 5.00 GΩ |
10 MΩ |
0.05 |
5.0 GΩ to 0.0 GΩ |
100 MΩ |
0.05 |
20.0 GΩ to 50.0 GΩ |
500 MΩ |
0.2 |
50 GΩ to 200 GΩ |
5 GΩ |
0.2 |
>200 GΩ |
Unspecified |
Unspecified |
2500 V |
<200 kΩ |
Unspecified |
Unspecified |
200 kΩ to 500 kΩ |
1 kΩ |
0.05 |
0.50 MΩ to 5.00 MΩ |
10 kΩ |
0.05 |
5.0 MΩ to 50.0 MΩ |
100 kΩ |
0.05 |
50 MΩ to 500 MΩ |
1 MΩ |
0.05 |
0.50 GΩ to 5.00 GΩ |
10 MΩ |
0.05 |
5.0 GΩ to 50.0 GΩ |
100 MΩ |
0.05 |
50 GΩ to 500 GΩ |
5 GΩ |
0.2 |
>500 GΩ |
Unspecified |
Unspecified |
Bar graph range: 0 Ω to 1 TΩ Insulation test voltage accuracy: -0 %, +10 % @ 1 mA load current Charging rate for capacitive load: 5 s/μF Discharging rate for capacitive load: 1.5 s/μF
|
Range |
Accuracy |
Leakage current measurement |
1 nA to 2 mA |
±(20 % + 2 nA) |
Capacitance measurement |
0.01 μF to 2.00 μF |
±(15 % rdg + 0.03 μF) |
Test voltage for insulation resistance |
250 V to 2500 V |
±(3 % + 3 V) |
Warning for energized circuit |
Warning range |
>30 V |
Timer (F1537 only) |
Range |
Resolution |
0 to 99 min |
Setting: 1 min |
Alert: 1 s |
Short circuit current |
F1535 |
>2mA |
F1537 |
>5mA |
ACV/DCV/resistance measurement (1537 only) |
Function |
Range |
Resolution |
Accuracy ±(% rdg + dgt) |
VAC |
0 V to 600.0 V |
0.1 V |
±(2 % +10) (45 Hz to 500 Hz) |
VDC |
0 V to 600.0 V |
0.1 V |
±(2 % +10) |
Resistance |
0Ω to 600.0Ω |
0.1Ω |
±(2 % +10) |
600Ω to 6000Ω |
1Ω |
6.00 kΩ to 60.00kΩ |
10Ω |